TY - JOUR
T1 - Evaluation of commercially available radioimmunoassays for the determination of luteinizing hormone in serum and urine. Application for the detection of endogenous rise of luteinizing hormone in an in vitro fertilization program
AU - Liu, Y.
AU - Naaktgeboren, N.
AU - Heip, J.
AU - Devroey, P.
AU - Temmerman, M.
AU - Verhoeven, N.
AU - Van Steirteghem, A. C.
PY - 1984
Y1 - 1984
N2 - Three commercially available RIA kits for serum and urinary LH were assessed for their usefulness to detect endogenous LH rise in patients receiving ovarian stimulation as part of an in vitro fertilizaton treatment for infertility. Prerequisites included a turn-around time of 5 hours for an assay of 100 tubes. The following parameters were evaluated: reproducibility of standard curve, sensitivity, precision profile, within- and between-assay precision, analytical drift, recovery and linearity. The Amerlex LH RIA kit was selected because of superior precision profiles, higher precision and better recovery tests.
AB - Three commercially available RIA kits for serum and urinary LH were assessed for their usefulness to detect endogenous LH rise in patients receiving ovarian stimulation as part of an in vitro fertilizaton treatment for infertility. Prerequisites included a turn-around time of 5 hours for an assay of 100 tubes. The following parameters were evaluated: reproducibility of standard curve, sensitivity, precision profile, within- and between-assay precision, analytical drift, recovery and linearity. The Amerlex LH RIA kit was selected because of superior precision profiles, higher precision and better recovery tests.
UR - http://www.scopus.com/inward/record.url?scp=0021710050&partnerID=8YFLogxK
M3 - Article
C2 - 6442546
AN - SCOPUS:0021710050
SN - 0003-3898
VL - 42
SP - 419
EP - 425
JO - Annales de Biologie Clinique
JF - Annales de Biologie Clinique
IS - 6
ER -