Abstract
Objective: To evaluate the performance of the point-of-care i-STAT device for creatinine measurement in Ugandan children with sickle cell anemia (SCA), and to determine the clinical impact of hydroxyurea-associated assay interference. Results description: This secondary analysis was nested within a randomized clinical trial involving 248 Ugandan children with SCA. The mean age at enrollment was 32 months, and 115/248 (46.2%) initiated hydroxyurea during 12 months of follow-up as part of routine clinical care. At enrollment, creatinine values measured by i-STAT and reference laboratory testing were clinically comparable. At follow-up, children receiving hydroxyurea had significantly higher creatinine values measured by i-STAT than by reference laboratory testing (mean difference + 0.19 mg/dL; p = 0.006), with several i-STAT results falling within the clinically abnormal range. No such discrepancy was observed among children not receiving hydroxyurea. In a subset of samples, serum hydroxyurea concentrations measured by high-performance liquid chromatography accounted for 73% of the variability in i-STAT creatinine values. These findings demonstrate clinically meaningful interference of hydroxyurea with enzymatic point-of-care creatinine testing, which may result in misclassification of kidney function and inappropriate clinical decision-making in children with SCA receiving hydroxyurea. ClinicalTrials.gov identifier: NCT03528434 (date registered: May 7th, 2018).
| Original language | English (US) |
|---|---|
| Article number | 338 |
| Journal | BMC Research Notes |
| Volume | 19 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - Dec 2026 |
| Externally published | Yes |
Keywords
- Creatinine
- Diagnostic accuracy
- Hydroxyurea
- Interference
- Point-of-care testing
- Sickle cell anemia
Fingerprint
Dive into the research topics of 'Hydroxyurea interferes with point-of-care creatinine testing in children with sickle cell anemia'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver